The following are the mechanical specifications for the color bar artwork for the X-Rite Eye-One iSis and iSis XL spectrophotometers. For more information about mechanical specifications, see the documentation that is provided with your spectrophotometer.
1 | First patch top edge |
2 | Leader edge |
3 | You can write in the first 1.2 cm (0.47 in.) |
4 | Clear space one |
5 | Bar height |
6 | Clear space two |
7 | Rows |
8 | Trailer edge |
9 | Columns |
10 | Positioning mark height |
11 | Positioning mark width |
12 | Chart width |
13 | Patch height |
14 | Patch width |
15 | Chart height |
16 | Positioning mark white space |
17 | First patch left edge |
Note: Before you can feed the chart of the color bar into the i1iSis spectrophotometer, you must cut the chart from the proof. Align the chart with the left side of the feed area on the iSis and iSis XL, and then feed the chart into the spectrophotometer.
Area | Dimension | Minimum | Maximum |
---|---|---|---|
Chart | Width | 6 cm (2.36 in.) | 23 cm (9.06 in.) XL: 33 cm (13 in.) |
Height | 17 cm (6.7 in. | 66 cm (26 in.) | |
Patch surrounding | Leader edge | 3.2 cm (1.26 in.) | 8 cm (3.15 in.) |
Trailer edge | 2.5 cm (0.98 in.) | ||
Clear space | 1 2 cm (0.79 in.) | ||
Clear space 2 | 6 mm (0.24 in.) | ||
Bar height | 5 mm (0.20 in.) | 8 mm (0.31 in.) | |
Positioning marks described below | |||
Patches | Patch array | 1 patch | 3000 patches |
Patch height | 6 mm (0.24 in.) | 20 mm (0.79 in.) | |
Patch width | 6 mm ( 0.24 in.) | 20 mm (0.79 in.) |
The positioning bar requirements are:
The positioning mark requirements are:
Dimension | Minimum |
---|---|
Positioning marker width | 5 mm (0.19 in.) |
Positioning marker height | 2.5 mm (0.09 in.) |
White space left of marker | Only 3 mm (0.11 in.) is supported |
White space right of marker | Only 3 mm (0.11 in.) is supported |