The following are the mechanical specifications for the color bar artwork for the X-Rite Eye-One iSis and iSis XL spectrophotometers. For more information about mechanical specifications, see the documentation that is provided with your spectrophotometer.

1First patch top edge
2Leader edge
3You can write in the first 1.2 cm (0.47 in.)
4Clear space one
5Bar height
6Clear space two
7Rows
8Trailer edge
9Columns
10Positioning mark height
11Positioning mark width
12Chart width
13Patch height
14Patch width
15Chart height
16Positioning mark white space
17First patch left edge

Note: Before you can feed the chart of the color bar into the i1iSis spectrophotometer, you must cut the chart from the proof. Align the chart with the left side of the feed area on the iSis and iSis XL, and then feed the chart into the spectrophotometer.

AreaDimensionMinimumMaximum
ChartWidth6 cm (2.36 in.)

23 cm (9.06 in.)

XL: 33 cm (13 in.)

Height 17 cm (6.7 in.66 cm (26 in.)
Patch surroundingLeader edge3.2 cm (1.26 in.)8 cm (3.15 in.)
Trailer edge2.5 cm (0.98 in.)
Clear space1 2 cm (0.79 in.)
Clear space 26 mm (0.24 in.)
Bar height5 mm (0.20 in.)8 mm (0.31 in.)
Positioning marks described below
PatchesPatch array1 patch3000 patches

Patch height6 mm (0.24 in.)20 mm (0.79 in.)

Patch width6 mm ( 0.24 in.)20 mm (0.79 in.)

The positioning bar requirements are:

The positioning mark requirements are:

DimensionMinimum
Positioning marker width 5 mm (0.19 in.)
Positioning marker height2.5 mm (0.09 in.)
White space left of markerOnly 3 mm (0.11 in.) is supported
White space right of markerOnly 3 mm (0.11 in.) is supported