Symptom

Causes and solutions

After you start calibration, the following message appears: Communication error - try unplugging and replugging EyeOne.

Confirm that the measuring device is connected. Firmly plug it into the computer.
Do not plug it into the USB connector on the keyboard, monitor, or USB hub.

If the message still appears, the measuring devices may be damaged. Try another measuring device.

After you start calibration, the following message appears: Calibration failed. Please check if the EyeOne is connected.

And the logs show: Calibration failed : New integration time: No light.

  1. Make sure that the shutter on the calibration plaque was open to expose the white tile when you "zero" the measuring device.

  2. Do not use Active USB Extension/Repeater cables when connecting the measuring device to the computer.

You start the calibration, but the monitor does not calibrate successfully, or one of the following messages appear:

  • Unable to setup monitor according to vendor instructions.
  • Error setting up display pixels for accurate color.
  1. Ensure that the monitor's image data cable and USB control cable are securely connected.

  2. Ensure that the measuring device cable is securely connected.

  3. Check the fixed luminance setting. The monitor may be unable to achieve the set value. If so, lower the setting and recalibrate.
    Note: Replace the monitor when it can no longer achieve a luminance level of 120.

  4. Ensure that you "zero" the measuring device on its calibration plate. Any light leakage will cause problems.
    Note: If you use the iOne Display 3, you do not need to perform the "zero" step.

  5. Check the measuring device for damage. If it is damaged, try a different measuring device.

You start the calibration using the iOne Display 3 colorimeter and the following message appears:

Calibration failed
New integration time: Too little light

To perform a successful calibration, make sure that the cover of the colorimeter is open.
After you start calibration, the following message appears: The black measurement has drifted too much - must recalibrate measurement device

This message might appear due to any of the following:

  • The EyeOne Pro colorimeter is not seated perfectly in the calibration plaque during the black calibration performed at the beginning.
  • The EyeOne Pro is older than Rev B or is malfunctioning.
  • Long waiting times have been encountered during the calibration process.
  • The colorimeter is stored in an environment that has a different temperature than where the monitor is located.
    Note: Always allow the colorimeter temperature to stabilize 10 minutes.
  • The colorimeter was not making complete or direct contact when performing the previous calibration.

 Solution:

  1. Quit the MVCalibrator application.
  2. At the Dock, click the Matchprint Virtual folder icon.
  3. Select MVCalibrator and in the MV Calibrate window that appears, on the right-hand side, click  Image .
  4. Make a note of the Fixed Luminance value.
  5. Switch to Max luminance. Click Save.
  6. Immediately switch back to the Fixed Luminance value. Click Save.
  7. Click Calibrate to start over.  

You start the calibration using the iOne Pro colorimeter and the following message appears:

White calculation: Sensor is saturated.

To perform a successful calibration, make sure that you open the protective slider until it clicks into place to access the white reference ceramic tile on the calibration plate.

You start calibration and the following message appears:

Monitor not allowing white point to be set.

If you are using a NEC UHD type monitor, the video signal might not be set properly.

Solution:

  1. Check the video level setting on the display and set to expand if necessary. 
  2. Verify that the DV Mode is standard and the color mode is Programmable
On systems with a NEC PA series monitor, the following message appears and stays for a while (possibly 50 seconds or more) when you switch between CALIBRATE and SETTING views.

Wait for display response...

For NEC PA series monitors, communication with the monitor takes more time than other models.

Solution: Wait until the process completes.

On system with large-screen NEC monitor (e.g. V554Q), the following message appears and cannot perform calibration.

Unable to setup monitor according to vendors instruction.

For such large-screen NEC monitors, the SpectraView Engine mode needs to be set to ON in order for Matchprint Virtual to control the monitor to perform calibration.

Solution: Via OSD (on-screen display) menu, set the SpectraView Engine mode to ON.

Table 2: Problems when performing monitor proofing

...